CHI-SQUARE PROTOCOL
REPETITIONS:
COUNTING TIME:
OK
>10
30 SEC
Figure 5-18
Input the number of repetitions desired (the minimum is 5 and the maximum is 30; at least 10
repetitions is recommended).
Input the counter time for each measurement (the minimum is 10 seconds and the maximum
is 9999 seconds).
Press
F4 [OK]
when the correct entries have been made.
CONVERSION FACTORS
Conversion Factors are constants used to convert measurements in counts per minute (cpm)
to disintegrations per minute (dpm). The CAPTUS
®
600 calculates dpm values by multiplying
the cpm measurement by the Conversion Factor of a particular nuclide. Conversion Factor
values are affected by geometry, detector efficiency, abundance of the gamma energy
peak(s) of the isotope, and the Region of Interest selected for that nuclide.
You can store Conversion Factors for all of the radionuclides already stored in your
CAPTUS
®
600 and for ten additional user-defined radionuclides, and for a “generic” nuclide.
Since some radiation safety regulations require you to document wipe test monitoring results
in dpm units, you should store Conversion Factors for the nuclides that could be present in a
wipe sample.
The CAPTUS
®
600 calculates Conversion Factors by measuring counts (in cpm) from a
calibrated sample of a specific radionuclide. Because the sample is calibrated, the CAPTUS
®
600 can calculate the dpm value of that sample. The CAPTUS
®
600 then divides the dpm
value by the cpm value to determine the Conversion Factor (CF) for that nuclide.
minute
per
counts
minute
per
tions
disintegra
=
de
radionucli
for
CF
November 07
SYSTEM INITIALIZATION
5-11
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......