System Power Requirements
CAUTION
If the input voltage to the following items is not within the stated limits,
the unit may not function correctly or may be damaged.
Line Voltage: 115 or 230 VAC depending on model ordered
Line Frequency: 60 or 50 Hz
Current:
0.7 Ampere @ 115V
0.4
Ampere
@
230V
Fuse: For 115V use T 3.15AH, 250V, (Time Delay)
For 230V use T 2.5AH, 250V, (Time Delay)
Line Filter:
Line filter is provided internally.
Use of a filtered line is recommended if excessive line noise is anticipated.
Environment Requirements
Operational
The instrument should be located where the level of the background radiation is as
low and as constant as possible.
The instrument should be located where the temperature is stable within a range of
+50°F to +85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-
condensing to warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of
27–31 inches of mercury (91–105 kilopascals).
Storage
The instrument should be stored where the temperature is stable and the range is
from +39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90%
non-condensing to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of
15–33 inches of mercury (51–112 kilopascals).
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
November 07
FUNCTIONAL & TECHNICAL DESCRIPTION
2-3
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......