CAPINTEC, INC.
CAPTUS
®
600
FUNCTIONAL DESCRIPTION
Thyroid Uptake
The CAPTUS
®
600 Thyroid Uptake / Well System is intended to be used by trained Nuclear
Medicine Technologists for performing Thyroid Uptake, Thyroid Bioassay, Wipe Test, Blood
Volume (with Cr-51 or I-125) Test, Shillings Test and RBC Survival Test.
Many of the Thyroid Uptake measurements are performed in exactly the same way at a
particular institution for a certain set of patients. To take advantage of this fact, the CAPTUS
®
600 Thyroid Uptake System is organized using the concept of protocols.
By using protocols, a set of measurement and data inputs can be defined in advance of
patient requirements. At the time of measurements, the system will prompt the technician for
the necessary patient data, dose information, dose administration times, and prompt the user
for which measurements are to be taken.
Upon taking the measurements, the CAPTUS
®
600 will perform the uptake calculations,
including any decay correction.
The basic concept of the Thyroid Uptake measurement with the CAPTUS
®
600 remains the
same for any protocol used. The uptake of the thyroid gland is equal to some portion of a
dose previously administered to the patient. However, there are three methods of calculation
to choose from.
Thyroid Uptake Nuclides
Thyroid Uptake Studies may be performed on the CAPTUS
®
600 using I123, I131 or Tc99m.
The nuclide is chosen when entering the protocol.
Pre-Dose Measurement
When a patient has been given the same nuclide that will be used in the Thyroid Uptake
Study, a more accurate result can be obtained by measuring the neck and thigh before the
dose is given. The result of this measurement will be subtracted from the actual neck and
thigh measurements.
The option to perform this Pre-Dose Measurement is chosen as part of the protocol.
Choosing this option does not force you to perform the measurement. It can be skipped if it is
not necessary.
Thyroid Uptake Protocols
The calculation method is chosen when the protocol is entered. The three calculation
methods follow.
November 07
FUNCTIONAL & TECHNICAL DESCRIPTION
2-7
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......