IFU-52/ REV H
9
Note: All status indicator lights will be green during normal operation.
NOTE:
VI. Front panel controls/connections
1.
On/Off push button
Turns main power on or off to the unit
2.
Water warmer On/Off
Push to turn water warmer on or off
3.
CO
2
Flow control
Controls flow rate of CO
2
output. The three levels are represented by three green indicators of “L”, “M”,
and “H” which are illuminated to indicate the flow rate setting. The ability to quickly turn off CO
2
flow is
available by pressing the flow control knob in. This will immediately turn the flow of CO
2
off and extinguish
the CO
2
flow light indicator. By turning the CO
2
flow control knob off and then back to your desired flow
rate setting (“L”, “M”, “H”) will restart the flow of CO
2
. The flow rates are as follows*:
Low – 1.4L/min
Medium – 2.4L/min
High – 3.5L/min
*
Flow rates shown are maximum free flow measurements at output of unit
Low gas pressure is indicated by the words “LOW PRESSURE” flashing to notify the user that the gas input
pressure has fallen below 25PSI.
4.
CO
2
gas output connector
Connect to tubing and GI endoscope.
Standard male luer lock connector.
5.
CO
2
gas heater On/Off
Push to toggle CO
2
gas heater On or Off
(The appearance of your device may differ slightly from picture shown above)
1
2
3
4
5
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 9
12/11/2015 9:31:56 AM