Certifications:
IEC-60601-1, IEC-60601-1-2, IEC-60601-2-18
Classification:
Class 1 Type B
IP rating (Ingress Protection):
IP24
Mechanical Specifications
Physical Dimensions:
Height
4 ¾” inches
121 mm
Width
7 ¾” inches
197 mm
Depth
13 ¾” inches
349 mm
Weight
10.5 pounds
4.8 kg
CO
2
Specifications
Free Flow rates (maximum unrestricted output to GI endoscope system):
Front panel setting
Low:
1.4 L/min
Medium:
2.4 L/min
High:
3.5 L/min
GI Endoscopic Flow rates (approximate flow rate at GI endoscope output):
Front panel setting
Low:
0.6-0.9 L/min*
Medium:
1.2-1.4 L/min*
High:
1.5-1.8 L/min*
*These values are approximate and are based on testing with respective models of GI endoscopes. User results
may vary depending on GI endoscope used, channel diameter, and channel length.
Output Connection:
Male luer lock
Input Connection:
¼” male flare fitting.
CO2 Heater:
20W, internal, flow through, with dual redundant temperature sensors.
CO2 Output temperature:
37°C (+3°C)
Operating Pressure:
Input Pressure
1900 PSI Maximum
25 PSI Minimum*
*The low pressure warning light will illuminate at 25 PSI or less, although the unit may still flow. It is
recommended to replenish the CO
2
supply when the light illuminates.
Pressure regulator set point
8 PSI Maximum
Internal safety relief valve
≤12 PSI
Grounding reliability can only be achieved when connected to a receptacle marked “Hospital Grade”.