IFU-52 / REV H
18
Max Output
Power
(Watts)
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/E1)(Sqrt P)
0.01
0.1
10
1
3.6893
7.3786
1.1666
2.3333
.3689
.7378
.1166
.2333
23.333
1.1666
100
11.6666
11.6666
.3689
3.6893
The EGA-501/EGA-501E is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the EGA-501/EGA-501E can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment and the EGA-501/EGA-501E as recommended below, according
to the maximum output power of the communications equipment.
Recommended Separations Distances for the EGA-501/EGA-501E
.1166
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Table 6 – Recommended Separation Distances
Between Portable And Mobile
RF Communications Equipment And The EGA-501/EGA-501E
EQUIPMENT and SYSTEMS That Are Not LIFE-SUPPORTING
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 18
12/11/2015 9:31:58 AM