Environmental requirements:
Operating Temperature:
16° to 24°C (+61° to 75°F)
Operating Relative Humidity:
30% to 75% non-condensing
Operating Pressure:
70 kPa-106 kPa_(10.2 PSI -15.4 PSI)
Storage Temperature:
0° to 40°C (32° to 104°F)
Storage Relative Humidity:
10% to 95% non-condensing
Storage Pressure:
70 kPa-106 kPa_(10.2 PSI -15.4 PSI)
Altitudes below 3000m
Accessory Items
The ENDO STRATUS
®
CO
2
insufflator is intended to be used only with the specific models
of compatible tubing sets, connectors and accessories identified in the table below and/ or
officially recommended by Medivators. Use of the device with accessories not identified in
the table below or officially recommended by Medivators may result in incompatibility and/ or
the risk of cross contamination and infection transmission.
Order Number
Description
100145CO2, 100145CO2U
ENDO SMARTCAP® tubing for OLYMPUS®140/240, 160/260,
180/280 (24 hr Use)
series GI Endoscopes
100150CO2, 100150CO2U
ENDO SMARTCAP tubing for OLYMPUS 100/200, 130/230 Series
(24 hr Use)
GI Endoscopes and PENTAX® GI Endoscopes
100165CO2, 100165CO2U
ENDO SMARTCAP tubing for FUJIFILM™ GI Endoscopes
(24 hr Use)
100141
ENDO SMARTCAP™ adaptor for FUJIFILM™ 530 Series
Colonoscopes (Single Use)
100605, 100605S
ENDOGATOR® hybrid tubing for OLYMPUS 140/160/180
(24 hr Use)
Series GI Endoscopes with OLYMPUS OFP or ENDOSTAT™ II Pump
Irrigation
Units
100606, 100606S
ENDOGATOR hybrid tubing for OLYMPUS 140/160/180
(24 hr Use)
Series GI Endoscopes with OLYMPUS AFU-100, OLYMPUS OFP or
EGP-100
irrigation
Units
100551
CO
2
Source Tubing
Replacement parts
Order Number
Description
EGA-7014
Power cord (E.U.)
EGP-042
Power cord (U.K.)
EGA-7011
High pressure hose
EGA-7012
Yoke
adapter
EGA-7010
Water bottle holder and warmer assembly
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 8
12/11/2015 9:31:56 AM