2. Installation
76
3-4) Press the exposure switch on the X-ray generator, and after the exposure, press the
END STUDY button. Transfer the SMPTE pattern image to the printer or the high-
definition monitor. [Fig. 2]
[Fig. 2]
3-5) Measure the densities of the 11 locations (0% to 100%) of test image grayscale on the
film or on the monitor. [Fig. 3]
* Measure the SMPTE image density on the film is measured using a densitometer.
Measure the SMPTE image density on the high-definition monitor using the
gradation analysis software.
* The data for the SMPTE test image grayscale transferred by the CXDI are the values
for the maximum density (3.20 in the case of the MLP 190) in the printer settings
which have been changed in 11 uniform steps.
[Fig. 3]