2. Installation
92
4-5) When Welcome to CCR appears, select “1. Set-Up...”.
4-6) When Setting Mode (0: Normal, 1: Expert) [0=0x0]: appears, select “0: Normal.”
4-7) When CCR SETUP MENU appears, select “7 Scan Sensor Setup”.
4-8) When Capture Device Configuration Table appears, enter the number of sensor to
which make the “Max Capture Device” recognized.
4-9) The dummies of “A/D Board Serial Number” are appeared: Enter the sensor serial
number for necessary type.
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
For example, when CXDI-50G is connected to Sensor 1, and CXDI-31 is connected to
Sensor 2:
Enter “
199
” for “A/D Board Serial Number for Sensor ID#1”
Enter “
0
” for “Custom Type”.
Enter “
3002
” for “A/D Board Serial Number for Sensor ID#2”
Enter “
6
” for “Custom Type”.
The CXDI software can now be launched on the control computer with the same
conditions in effect as if an imaging unit were connected.
@@@@@@@@ Capture Device Configuration Table @@@@@@@@
Max Capture Devices [4 = 0x4] : 4
Å
Number of imaging units connected
@@@@@@@@ Capture Device Configuration No.0 (SensorID#1 OPU)@@@@@@@@
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x123
------A/D Board Serial Number 0-3 -> 40G : 0x123
A/D Board Serial Number for SensorID#1 [0x199 = 409] :
Custom Type[0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE 100um
5:CASSETTE 14X17_160um 6:CASSETTE 9X11_160um] [0 = 0x0] : 0
Field of View Rotation (0:No 1:Yes) [0= 0x0] : 0
Constant for Exposure Index [0.000000] : 0.000000
@@@@@@@@ Capture Device Configuration No.1 (SensorID#2 OPU)@@@@@@@@
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
A/D Board Serial Number for SensorID#2 [0x2009 = 8201]
Custom Type[0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE 100um
5:CASSETTE 14X17_160um 6:CASSETTE 9X11_160um] [0 = 0x0] : 0
Field of View Rotation (0:No 1:Yes) [0= 0x0] : 0
Constant for Exposure Index [-1.000000] : -1.000000
@@@@@@@@ Capture Device Configuration No.2 (SensorID#3 OPU)@@@@@@@@
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
A/D Board Serial Number for SensorID#3 [0x3002 = 12290]
Custom Type[0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE 100um
5:CASSETTE 14X17_160um 6:CASSETTE 9X11_160um] [0 = 0x0] : 0
Field of View Rotation (0:No 1:Yes) [0= 0x0] : 0
Constant for Exposure Index [-1.000000] : -1.000000
Dummy Serial No. for
Sensor 1
Dummy Serial No. for
Sensor 2
Dummy Serial No. for
Sensor 3
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...