![Canon CXDI-60G Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/canon/cxdi-60g/cxdi-60g_service-manual_4009843078.webp)
2. Installation
62
* Both the -p and -v
parameters should be used.
These settings override the -
s parameter.
-V filename
This parameter dumps the
transfer log.
* The parameter is used for
analysis after the transfer
software saves the
debugging character string
displayed on the console to a
file with a designated
filename.
It is used only when
problems occur.
* This parameter does not affect DICOM
data transfer.
-I
A-RELEASE-RES is
ignored.
* This parameter is used
simply as “-I”
* This is used when the error message
[130012 Peer aborted Association (or
never connected)]occurs even though the
DICOM data transfer was successful.
Æ
This is used based on the connected
printers.
-jn
This is the time to take
timeout.
* Sets the time to take timeout
in seconds.
* This parameter is to be changed when
taking timeout.
-k
level = 0
1
2
* The -k option has been
implemented to normalize
DICOM headers
Note
The parameters -p, -v, and -V filename display the log on the console. Therefore, be sure
to always erase these parameters before operation by the user.
In the past, -k2 was used to fix troubles, but it has now been set as the default has been changed
to -k2 to ensure precise compliance with the DICOM code. At sites wishing to adhere to their
existing window values, -k1 should be used.
level: 0:Do not delete DICOM tags. (Transfer all headers similar to storage.)
1:Delete Groups 0008, 0010, 0018, 0019 and 0020.
2:Delete Groups 0008, 0010, 0018, 0019, 0020 and Elements (0028,0030), (0028,1050),
(0028,1051), (0028,1052), (0028,1053), (0028,1054). Default
Image Pixel Spacing (0028,0030)
Window Center (0028,1050)
Window Width (0028,1051)
Rescale Intercept (0028,1052)
Rescale slope (0028,1053)
Rescale Type (0028,1054)
For groups 0008, 0010, 0018, 0019, 0020, refer to DICOM manifesto.
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...