2. Installation
86
(13) Changing the Total Image Count
1) Purpose
When the imaging unit is replaced (including the replacement of the LANMIT) for
servicing, the total image count displayed on the user screen can be returned to “0”if
necessary.
2) Notes
2-1) The CXDI is connected by the system.
2-2) Set the CXDI software so that it does not start up.
2-3) Files are overwritten, and so be careful when performing the procedure.
2-4) As an extra precaution, write down the numerical values before overwriting them.
2-5) The overwritten counter becomes valid the next time that the CXDI software is started
up.
3) Procedure
3-1) Turn on the control computer, and then start up Windows XP.
3-2) Right click on My Computer on the Desktop screen of Windows XP, and open
Explorer.
3-3) A file called “ExpResult.ini” is contained in the CCR folder.
Open this file. (D:\CCR\ExpResult.ini) [Fig 1]
[Fig 1]
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...