![Canon CXDI-60G Скачать руководство пользователя страница 176](http://html1.mh-extra.com/html/canon/cxdi-60g/cxdi-60g_service-manual_4009843176.webp)
6. Troubleshooting
12
5. Gathering Information About the Problem
It is recommended to replace the whole unit (imaging unit, power box) if any problem occurs at
the customer’s site in order to reduce the downtime of the system.
It is desirable to have the bad unit brought back and repaired at the office of the distributor
(agent). (Although it depends on the kind of the problem.)
This section explains about the information (log file) required to know what has occurred and
what to be done.
[Probable problems]
-
System connection failure and condition setting failure during installation.
-
Electrical and mechanical system failure, malfunctioning, bad image, noise, and
communication error
-
Software and specifications problem
[Required Information]
*Required log file
Execute the log collection tool to save the following files.
(1) Environmental information (such as version of the CXDI software, composition of hardware,
optional software and the versions of software)
(2) Dr. Watson log
C:\Documents and Settings\All Users\Application
Data\Microsoft\Drwatson\drwts32.log
(3) Event viewer system log
C:\WINDOWS\system32\config\SysEvent.Evt
(4) The whole D:\Ccr folder
Reference:
For details on the procedure for gathering the files when image problems occurs, refer to
SIDR-08-003.
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...