2. Installation
33
3.2 Network setting for Screwcap.ini
The CXDI software communicates with the imaging part through Screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
imaging unit such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Since the Screwcap.ini factory setting is the following default setting (see the figure below), it is
not necessary to edit the setting unless you changed the network protocol TCP/IP setting for the
LAN card that communicates with the imaging unit in the previous item, “3.1 TCP/IP setting for
the control computer network card”.
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
*
1
Imaging Unit IP address:
This address must be the same as the IP address stored in
the next item, “3.3 Network setting stored in the sensor”.
[HostInfo]
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=
1
[Sensor1]
IpAddress=
192.168.100.11
CommandPort=12121
DataPort=12122
[Sensor2]
IpAddress=192.168.100.12
CommandPort=12121
DataPort=12122
[Sensor3]
IpAddress=192.168.100.13
CommandPort=12121
DataPort=12122
[Sensor4]
IpAddress=192.168.100.14
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
[SensorInfo]
SensorNum=1 Confirm this
Sensor IP address.
*1
This must be the same as the
IP address stored in the
imaging part.
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...