20
4)
While pressing the protuberance on the lock base (6) against the indentation in the
slide switch (5), and fasten them together securely using the screw (XA4-1200-609).
Now tighten up the screw (XA4-1200-609) which was temporarily secured in step 3).
* Bear in mind that using excessive torque for tightening the screws will damage
the slide switch (6).
* Be careful when tightening up the two screws so as to avoid damaging their
heads.
* Reference value for the tightening torque: 1kgf·cm (9.8N·cm)
Press the protuberance
against this part.
Fasten together securely
using the screw
(5)
(6)
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...