4
2 CXDI SYSTEM II
The CXDI SYSTEM II unit consists of the power box, cable with AC plug, X-ray interface
cable and remote switch.
2.1 Power box
This consists of the 60 sensor power supply, PWB-60XRAY board and X-ray power cable
assembly, and it incorporates the functions for transferring signals between the control computer
and imaging unit, supplying power to the imaging unit and relaying the signal transfer to and
from the X-ray generator. One imaging unit can be connected to this board: It is not possible to
connect a multiple number of imaging units. The sensor can be replaced with the other one
which supports the power box by connecting the connectors of the sensor cable S150-60
(imaging unit side) and sensor cable P630 (power box side). Before the sensor is replaced, the
power of the power box must be turned off. Trouble may develop in the system if the sensor is
replaced with the power on.
To imaging unit
To X-ray generator
To control PC
Remote switch
Mainly signal systems
Power supply system
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...