2. Installation
53
4-2) Storage settings
A) Open the output destination setting dialog from the user menu.
System
Æ
SETUP MENU
Æ
DESTINATION
Æ
STORAGE
* Up to four storage units (2 of which can be used for output at the same
time) can be set.
B) Press the “Storage1” button, and input the following items based on the pre-
installation investigation details.
a. Storage host name (IP address) b. Port number c. Transmission destination
title
[Fig 4]
C)
Press the “SET” button, and input the parameters.
(Normally, the parameters need not be set. They must be input only when the
need arises.)
* A space delimiter must be input between each of the parameters.
[Fig 5]
a
b
c
Input parameters
as needed
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...