4. Repair Guide
4
2.3 Replacing the sensor cable
1) Disconnect the connector of the sensor cable which is connected to the J2 connector on the
LED board.
Do not pull the wires to disconnect the connector. Using needle-nosed pliers, push the
three positions uniformly indicated by the red arrows in the picture below. Be careful
when removing the connector so as to avoid damaging the connector.
After disconnecting the connector, check that the IC soldered to the LED board is not
damaged.
2) Disconnect the connector of the sensor cable which is connected to the J1 connector on the
LED board.
Do not pull the wires to disconnect the connector. Using needle-nosed pliers, push the
three positions uniformly indicated by the red arrows in the picture below. Be careful
when removing the connector so as to avoid damaging the connector.
After disconnecting the connector, check that the IC soldered to the LED board is not
damaged.
Содержание CXDI-60G
Страница 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Страница 43: ...2 Installation 27 During normal imaging During timeout due to RX_REQ not negating Timeout 1100msec ...
Страница 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 116: ...2 Installation Manual 100 3 Remote Switch Mass 0 9 Kg inc cable Unit mm Dimension tolerance 1 Scale 1 N ...
Страница 120: ...3 Function 1 1 Imaging Unit 1 1 Block Diagram Imaging Unit Power Box Mainly signal systems Power supply system ...
Страница 154: ...CXDI SYSTEM Rev 01 CONTENTS IMAGING UNIT 1 POWER BOX 2 GRID UNID 3 SCHEMATIC DIAGRAM PWB 60 X RAY I F 4 ...
Страница 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Страница 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...