CANNON Instrument Company® | Getting Started
27
3.
Click
Verify Known KV
in the
Methods
screen. The field entry screen open. Refer to Figure 36.
Figure 36 — Methods Screen
4.
Enter the
Sample ID
. Then click the green checkmark to submit the entry. Refer to Figure 37.
Another field entry screen opens for the viscosity value.
Figure 37 — Sample ID Entry Field
5.
Enter the viscosity value (in cSt) of the test oil. Then click the green checkmark to submit the entry
and return to the Main Menu screen.
6.
Click the
Run
button.
7.
When the test is complete, select the
Results
button to review the sample test results.
Review and Export Data (
Touchscreen
)
Once a test completes, the results can be exported to a USB drive in CSV format.
1.
From the main screen, click the
Results
button.
2.
Review the results.
Verify Known KV option
in the Methods Screen