![Cannon CAV 4.1 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/cannon/cav-4-1/cav-4-1_operation-manual_3930099024.webp)
18
| CAV 4.1 Operation Manual
1.
If raised, lower the display hood and plumbing lid.
2.
Carefully insert a thermometer through the access hole in the display hood and into the bath until
the tip of the thermometer is in the middle of the capillary portion of the viscometer tube. The
thermometer passes through the same hole in the top of the bath that is used to fill the unit with
fluid. Refer to Figure 19.
Figure 19 — Thermometer Access Hole
3.
Wait for the thermometer to adjust to the temperature of the bath fluid. Read the reference
temperature.
4.
From the main screen, click the current bath temperature button
for the bath. This opens
the
Temperature
Calibration
screen, as shown in Figure 20.
Figure 20 — Temperature Calibration Screen
5.
Click the
Reference Temperature
field. A field entry screen opens.
Thermometer
access hole