Section 7. Installation
264
If you are doing a dry run, assume the temperature is 10 °C.
a. Enter
CRBasic EXAMPLE: PT100 BrHalf4W() Four-Wire Half-Bridge
Measurement
(p. 265)
into the CR800. It is already programmed with the
excitation voltage from step 3 and RS
0
from step 4.
b. Place PT100 in medium to measure.
c. Measure with
BrHalf4W()
. If you are doing a dry run, assume the result
of
Resistance()
= X
10
= 1.039.
d. Calculate RS
10
:
RS
10
= X
10
• Rf = 103900
6. Calculate RS
10
/RS
0
, K, and temperature:
a. RS
10
/RS
0
= 1.039
b. K = (RS
10
/RS
0
)-
1 = 0.039
c. T = g * K^4 + h * K^3 + i * K^2 + j * K = 9.99 °C
d. T = (SQRT(d * (RS
10
/RS
0
) + e) -
a) / f = 9.99 °C
1
A Campbell Scientific terminal-input module (TIM) can be used to complete the resistive bridge
circuit. Refer to the appendix
Passive-Signal Conditioners — List
(p. 563).
2
The magnitude of the excitation voltage does not matter in mathematical terms because the result of
the measurement is a ratio rather than an absolute magnitude, but it does matter in terms of reducing
the effect of electromagnetic noise and of losing of resolution. A maximum excitation helps drown
out noise. A minimum input-voltage range helps preserve resolution.
3
Get this value from a PRT resistance-to-temperature table
CRBasic Programs and Notes
PT100 BrHalf4W() Four-Wire Half-Bridge Calibration
'This program example demonstrates the calibration of a 100-ohm PRT (PT100) in a four-wire
'half bridge with voltage excitation. See adjacent procedure and schematic.
'Declare constants and variables:
Const
Rf = 100000
'Value of bridge resistor
Public
X
'Raw output from the bridge
Public
RS0
'Calculated PT100 resistance at 0 °C
BeginProg
Scan
(1,Sec,0,0)
....'Measure X:
'BrHalf4W(Dest,Reps,Range1,Range2,DiffChan,ExChan,MeasPEx,ExmV,RevEx,RevDiff,
' SettlingTime,Integ,Mult,Offset)
BrHalf4W
(X,1,mV25,mV25,1,Vx1,1,2053,True,True,0,250,1,0)
'Calculate RS0:
RS0 = X * Rf
NextScan
EndProg
Содержание CR800 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 34: ......
Страница 54: ......
Страница 92: ......
Страница 190: ...Section 7 Installation 190 FIGURE 40 Running Average Frequency Response FIGURE 41 Running Average Signal Attenuation ...
Страница 310: ......
Страница 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Страница 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Страница 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Страница 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Страница 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Страница 456: ......
Страница 462: ......
Страница 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Страница 524: ......
Страница 526: ......
Страница 556: ......
Страница 558: ......
Страница 560: ......
Страница 597: ......