Section 10. Troubleshooting
467
10.5.3.1.1 Voltage Measurements
The CR800 has the following user-selectable voltage ranges: ±5000 mV, ±2500
mV, ±250 mV, and ±25 mV. Input signals that exceed these ranges result in an
over-range indicated by a
NAN
for the measured result. With auto range to
automatically select the best input range, a
NAN
indicates that either one or both
of the two measurements in the auto-range sequence over ranged. See
Troubleshooting — Auto Self-Calibration Errors
.
A voltage input not connected to a sensor is floating and the resulting measured
voltage often remains near the voltage of the previous measurement. Floating
measurements tend to wander in time, and can mimic a valid measurement. The
C
(open input detect/common-mode null) range-code option is used to force a
NAN result for open (floating) inputs.
10.5.3.1.2 SDI-12 Measurements
NAN
is loaded into the first
SDI12Recorder()
variable under the following
conditions:
•
CR800 is busy with terminal commands
•
When the command is an invalid command.
•
When the sensor aborts with CR LF and there is no data.
•
When
0
is returned for the number of values in response to the
M!
or
C!
command.
10.5.3.2 Floating-Point Math, NAN, and ±INF
Related Topics:
•
Floating-Point Arithmetic
(p. 160)
•
Floating-Point Math, NAN, and ±INF
(p. 467)
•
TABLE: Data Types in Variable Memory
(p. 127)
Table
Math Expressions and CRBasic Results
(p. 468)
lists math expressions, their
CRBasic form, and IEEE floating point-math result loaded into variables declared
as FLOAT or STRING.
10.5.3.3 Data Types, NAN, and ±INF
NAN
and
±INF
are presented differently depending on the declared-variable data
type. Further, they are recorded differently depending on the final-memory data
type chosen compounded with the declared-variable data type used as the source
(
TABLE: Variable and FS Data Types with NAN and ±INF
(p. 468)
). For example,
INF
, in a variable declared
As LONG
, is represented by the integer –
2147483648
. When that variable is used as the source, the final-memory word
when sampled as UINT2 is stored as 0.
Содержание CR800 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 34: ......
Страница 54: ......
Страница 92: ......
Страница 190: ...Section 7 Installation 190 FIGURE 40 Running Average Frequency Response FIGURE 41 Running Average Signal Attenuation ...
Страница 310: ......
Страница 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Страница 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Страница 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Страница 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Страница 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Страница 456: ......
Страница 462: ......
Страница 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Страница 524: ......
Страница 526: ......
Страница 556: ......
Страница 558: ......
Страница 560: ......
Страница 597: ......