Model 4420-V001
DIGISTANT
®
Page 111
SOURce:VOLTage:DELTa
DESCRIPTION:
This instruction is meant to set a delta-X voltage value.
SYNTAX:
SOURce:VOLTage:DELTa <parameter>
Parameters:
Float value, with optional unit.
Value range:
0 to 22MA
Valid units:
UV, MV, V, KV, MAV
Default unit:
V
REMARKS:
Enquiry form:
SOURce:VOLTage:DELTa?
On this enquiry, the delta-X voltage value is supplied in exponential
form (as volts) without an explicit unit.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:TCOuple:DELTa
DESCRIPTION:This instruction is meant to set a delta-X TC-value.
SYNTAX:
SOURce:TCOuple:DELTa <parameter>
Parameters:
Float value, with optional unit.
Value range:
0 to 3000 K
Valid units:
C or CEL, F or FAR and K
Default unit:
Depending on the setting.
REMARKS:
Enquiry form:
SOURce:TCOuple:DELTa?
On this enquiry, the actual delta-X value is supplied as a numerical
value followed by the currently selected unit.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:RESistance:DELTa
DESCRIPTION:
This instruction is meant to set a delta-X RTD-value.
SYNTAX:
SOURce:RESistance:DELTa <parameter>
Parameters:
Float value, with optional unit.
Value range:
0 to 3000 K or 0 to 2kΩ
Valid units:
C or CEL, F or FAR, K, UOHM,
MOHM, OHM, KOHM, MAOHM
Default unit:
Depending on the setting.
REMARKS:
Enquiry form:
SOURce:RESistance:DELTa?
On this enquiry, the actual delta-X RTD-value is supplied as a
numerical value followed by the currently selected unit.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode