Model 4420-V001
DIGISTANT
®
Page 109
SOURce:TCOuple[:LEVel][:IMMediate][:AMPLitude]
DESCRIPTION:
This instruction is meant to simulate a temperature for a thermocouple.
SYNTAX:
SOURce:TCOuple[:LEVel][:IMMediate][:AMPLitude] <parameter>
Parameters:
Float value, with optional unit.
Value range:
Depending on the thermocouple
Valid units:
C or CEL, F or FAR and K
Default unit:
Depending on the setting.
REMARKS:
Enquiry form:
SOURce:TCOuple[:LEVel][:IMMediate][:AMPLitude]]?
On this enquiry, the actual temperature transmission value is
supplied as a numerical value followed by the currently selected
unit.
Restrictions:
Instruction not permissible if - device in calibration mode,
- measurement in progress,
- device in ramp mode.
SOURce:RESistance[:LEVel][:IMMediate][:AMPLitude]
DESCRIPTION:
This instruction is meant to simulate a temperature or resistance
value for a Pt 100 or Ni 100.
SYNTAX: SOURce:RESistance[:LEVel][:IMMediate][:AMPLitude]
<parameter>
Parameters:
Float value, with optional unit.
Value range: Depending on the RTD
Valid units:
C or CEL, F or FAR, K, UOHM, MOHM, OHM, KOHM, MAOHM
Default unit: Depending on the setting.
If a changeover is made between OHM and a temperature
unit, it becomes the new default unit.
REMARKS:
Enquiry form:
SOURce:RESistance[:LEVel][:IMMediate][:AMPLitude]?
On this enquiry, the actual RTD transmission value is supplied as a
numerical value followed by the currently selected unit.
Restrictions:
Instruction not permissible if - device in calibration mode,
- (only) measurement in progress,
- device in ramp mode.
SOURce:STORe
DESCRIPTION:
The actual transmission value and corresponding delta-X value are stored.
SYNTAX:
SOURce:STORe <parameter>
Parameters:
Integer value between 0 and 9. This number corresponds to the
memory number under which the value pair is stored.
REMARKS:
Enquiry form:
No enquiry form
Restrictions:
Instruction not permissible if - device in calibration mode,
- (only) measurement in progress,
- device in ramp mode.