Model 4420-V001
DIGISTANT
®
Page 132
*WA
Ι
DESCRIPTION:
This instruction sets a device such that all instructions are executed sequentially.
This instruction does not have any effect in the 6706, as
instructions are always processed sequentially.
SYNTAX:
*WA
Ι
Parameters:
No parameters
REMARKS:
Enquiry form:
No enquiry form
*SAV
DESCRIPTION:
A device setting is stored.
Up to 4 settings can be stored.
SYNTAX:
*SAV <parameter>
Parameters:
A number under which the device setting is to be stored.
Value range: 1 <=number<= 4
REMARKS:
Enquiry form:
No enquiry form
Response: -
Restrictions:
Instruction not permissible if - device in ramp mode
- device in calibration mode
- measurement in progress
*RCL
DESCRIPTION:
A stored device setting is loaded.
Up to 5 settings can be loaded.
SYNTAX:
*RCL <parameter>
Parameters:
The number of the device setting to be loaded.
Value range: 1 <= number <= 5
5 loads the default setting.
REMARKS:
Enquiry form:
No enquiry form
Restrictions:
Instruction not permissible if - device in ramp mode
- device in calibration mode
- measurement in progress