Model 4420-V001
DIGISTANT
®
Page 121
SOURce:LIST:WAVeform
DESCRIPTION:
The waveform for ramp 2 is set here.
SYNTAX:
SOURce:LIST:WAVeform <parameter>
Parameters:
TRIangle ->
Triangular waveform.
SAWTooth ->
Sawtooth waveform.
REMARKS:
Enquiry form:
SOURce:LIST:WAVeform?
Parameters:
TRI
->
If a triangular waveform has been set.
SAWT
->
If a sawtooth waveform has been set.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:COUNt
DESCRIPTION:
The number of passes by ramp 2 is set here.
SYNTAX:
SOURce:LIST:COUNt <parameter>
Parameters:
Number of passes.
Value range: 0 to 99
0 -> Continuous pass
REMARKS:
Enquiry form:
SOURce:LIST:COUNt?
Response:
Set number of passes.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:POINts
DESCRIPTION:
The number of active stages for ramp 2 is set.
SYNTAX:
SOURce:LIST:POINts <parameter>
Parameters:
Number of active stages.
Value range: 2 to 30
REMARKS:
Enquiry form:
SOURce:LIST:POINts?
Response:
The current number of stages for ramp 2.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:NUMBer
DESCRIPTION:
A list number is sent from ramp 2 to the device. This list contains
a total of 30 value-pairs (transmission values and delta-time). The
value-pair corresponding to this list number can be set via the
interface.
SYNTAX:
SOURce:LIST:NUMBer <parameter>
Parameters:
The list number
Value range: 1 to 30
REMARKS:
Enquiry form:
SOURce:LIST:NUMBer?
Содержание DIGISTANT 4420-V001
Страница 4: ...Model 4420 V001 DIGISTANT Page 4...
Страница 12: ...Model 4420 V001 DIGISTANT Page 12...
Страница 20: ...Model 4420 V001 DIGISTANT Page 20...
Страница 145: ...Model 4420 V001 DIGISTANT Page 145 7 1 Measure and source 7 Technical data...