Model 4420-V001
DIGISTANT
®
Page 67
This menu is obtained by pressing the
FCT
and
RTD
keys.
The LCD provides the following display, for example:
Various parameters for RTD measurement and transmission are set in this menu.
The following parameters can be selected with the
▲
and
▼
cursor keys.
RTD selection:
The
→
and
←
keys are used to toggle between the following RTD types:
Pt100
(Pt100 is selected)
Ni100
(Ni100 is selected)
Pt200
(Pt200 is selected)
Pt500
(Pt500 is selected)
Pt1000
(Pt1000 is selected)
RL-OUT setting:
Here, you can specify whether or not the lead resistance RL should be considered for RTD
transmission.
The
→
and
←
keys can be used to toggle between the following options:
ON
RL is subtracted from the RTD measurement value
OFF
RL is not considered.
RL entry:
The lead resistance (for RTD transmission and measurement) can be adjusted over a value
range of 1 Ω bis 99 Ω with the
0
9 number keys.
Press the ENT key to save the new RTD setting and return to the measurement or
transmission mode.
4.16 RTD setting
R T D - S E T T I N G
R T D : P T 1 0 0