Model 4420-V001
DIGISTANT
®
Page 122
Response:
The currently selected list number.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:CURRent
DESCRIPTION:
For setting ramp 2. The current value of the list element
corresponding to the selected list number is set here. The list
number is selected with the SOURce:LIST:NUMBer instruction.
SYNTAX:
SOURce:LIST:CURRent <parameter>
Parameters:
Numerical value, with optional unit.
Value range:
-22 to 22mA
Valid units:
UA, MA, A, KA, MAA
Default unit:
A
REMARKS:
Enquiry form:
SOURce:LIST:CURRent?
Response:
The set current value (in amperes) for the selected list element.
The unit string is also transmitted.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:VOLTage
DESCRIPTION:
For setting ramp 2. The voltage value of the list element
corresponding to the selected list number is set here. The list
number is selected with the SOURce:LIST:NUMBer instruction.
SYNTAX:
SOURce:LIST:VOLTage <parameter>
Parameters:
Numerical value, with optional unit.
Value range:
-11 to 11V
Valid units:
UV, MV, V, KV, MAV
Default unit:
V
REMARKS:
Enquiry form:
SOURce:LIST:VOLTage?
Response:
The set voltage value (in volts) for the selected list element.
The unit string is also transmitted.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode
SOURce:LIST:TCOuple
DESCRIPTION:
For setting ramp 2. The TC value of the list element corresponding
to the selected list number is set here. The list number is selected
with the SOURce:LIST:NUMBer instruction.
SYNTAX:
SOURce:LIST:TCOuple <parameter>
Parameters:
Numerical value, with optional unit.
Value range:
0 to 3000 K or 0 to 2kΩ
Valid units:
C or CEL, F or FAR and K.
Default unit:
Depending on the setting
REMARKS:
Enquiry form:
SOURce:LIST:TCOuple?
Response:
The set TC value for the selected list element with the selected unit.
The unit string is also transmitted.
Restrictions:
Instruction not permissible if - device in calibration mode
- (only) measurement in progress
- device in ramp mode