Model 4420-V001
DIGISTANT
®
Page 30
4.2.4 Temperature measurement with a Pt100/200/500/1000
and Ni100
Connect the resistance sensor in 4 wire configuration to a 6-pin Lemo-plug (size 1).
ext.
Pt100
DIGISTANT
®
4420-V001
2
6
3
4
5
FE
PF
+ V
+ I
-
V
-
I
Ni100
Pt100
Pt200
Pt500
Pt1000
-
Switch on the device and connect it as shown above.
RTD setting
-
This menu is obtained by selecting the [FCT] and [RTD] functions.
The following display appears on the LCD:
-
Various parameters for RTD measurement and transmission are set in this menu.
-
The different parameters are selected with the
▲
and
▼
cursor keys.
RTD selection
-
The
→
and
←
keys are used to toggle between the following RTD types:
Pt100 (Pt100 is selected)
Ni100 (Ni100 is selected)
Pt200 (Pt200 is selected)
Pt500 (Pt200 is selected)
Pt1000 (Pt1000 is selected)
RL-OUT setting
-
No setting is required for measurement.
R T D - S E T T I N G
R T D :
1