Model 4420-V001
DIGISTANT
®
Page 108
SOURce:CORRection:OFFSet:STATe
DESCRIPTION:
Here, you can specify whether or not the lead resistance RL should
be considered for RTD transmission.
SYNTAX:
SOURce:CORRection:OFFSet:STATe <parameter>
Parameters:
ON or 1 -> RL is added to the RTD transmission value,
OFF or 0 -> RL is not considered.
REMARKS:
Enquiry form:
SOURce:CORRection:OFFSet:STATe?
Response:
1: -> If the RL function for RTD transmission is active,
0: -> If the RL function for RTD transmission is inactive.
Restrictions:
Instruction not permissible if - device in calibration mode,
- measurement in progress,
- device in ramp mode.
SOURce:CORRent[:LEVel][:IMMediate][:AMPLitude]
DESCRIPTION:
This instruction is meant to output a current value.
SYNTAX:
SOURce:CURRent[:LEVel][:IMMediate][:AMPLitude] <parameter>
Parameters:
Float value, with optional unit.
Value range: - 22MA to 22MA
Valid units:
UA, MA, A, KA, MAA
Default unit:
A
REMARKS:
Enquiry form:
SOURce:CURRent[:LEVel][:IMMediate][:AMPLitude]?
On this enquiry, the actual current transmission value is supplied in
exponential form (as amperes) without an explicit unit.
Restrictions:
Instruction not permissible if - device in calibration mode,
- (only) measurement in progress,
- device in ramp mode.
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]
DESCRIPTION:
This instruction is meant to output a voltage value.
SYNTAX:
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] <parameter>
Parameters:
Float value, with optional unit.
Value range:
Depending on the selection
Valid units:
UV, MV, V, KV, MAV
Default unit:
V
REMARKS:
Enquiry form:
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
On this enquiry, the actual voltage transmission value is supplied in
exponential form (as volts) without an explicit unit.
Restrictions:
Instruction not permissible if - device in calibration mode,
- (only) measurement in progress,
- device in ramp mode.