133
of 216
Parameters in the menu "Setup realtime switchpoints S1 to S6" (menu 12)
Value
Limit threshold
Enter the limit threshold
Units
Units display
Displays the units for the measurement
channel
Channel
X, Y
Select the active measurement channel for the
respective switching signals S1 to S6
Active
High / Low
Select whether signal is active high or active
low
Reference
Absolute / Y trigger
Reference for measurement channel X
Absolute:
The limit threshold is referred to the absolute
value of measurement channel X.
Y trigger:
The limit threshold is referred to the trigger
event (for further details, please see section
6.3.2.2 "Measurement curve reference" on
page 107).
Note:
You can only select the Y trigger reference if the “Reference” parameter in the “Measurement
mode” menu is set to Y-Trigger (overrun or underrun).
1
In measurement mode, tap anywhere on the touchscreen. The icon appears in the
bottom-right corner.
2
Tap to open the "Configuration Main Menu".
3
Tap the "Program Setup" icon.
4
Tap the "Switchpoints" icon.
5
The two-page menu that opens outlines in
red those fields that can be edited for the
switching points S1 to S6. To edit the setting,
tap the relevant field and hence change the
selected setting. You can also use the
keypad to enter values.
Note:
You can only select the Y trigger
reference if crossing the Y trigger
threshold (Y trigger >> or Y trigger
<<) is set as the "Reference"
parameter in the "Measurement
mode" menu.
Note:
Outputs S1 and S2 are permanently assigned to specific PLC outputs. The assignment for
outputs S3 to S6 can be custom-set (see section 6.1.2 "PLC outputs" on page 46).