280
SAW sensor – Maintenance
Type 8098 FLOWave L
To calibrate the offset value of the acoustic transmission factor, do the following:
→
Make sure the liquid in the pipe is the liquid to be measured.
→
Charge the pipe. To avoid bubbles and air in the pipe, make sure it is full of liquid.
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Add. meas. values
→
Acoustic transmission factor
→
Offset
→
Teach-in by reference
The
Current settings
are displayed.
→
Start the calibration of the offset value.
→
After 30 s, enter the acoustic transmission factor of the reference liquid.
The
New settings
are displayed.
→
Save.
The offset value of the acoustic transmission factor is calibrated.
If the calibration fails, a message is displayed. Refer to chapter 24.10 Messages due to calibration or
→
Acknowledge the message to go back to the parent menu.
English
Содержание Flowave 8098
Страница 1: ...Operating Instructions Type 8098 FLOWave L SAW Flowmeter Software version A 04 00 00 and higher...
Страница 4: ...General contents 4 Type 8098 FLOWave L English...
Страница 12: ...12 Type 8098 FLOWave L English...
Страница 36: ...36 Technical data Type 8098 FLOWave L English...
Страница 72: ...72 Type 8098 FLOWave L English...
Страница 78: ...78 Commissioning Type 8098 FLOWave L English...
Страница 96: ...96 Type 8098 FLOWave L English...
Страница 144: ...144 Type 8098 FLOWave L English...
Страница 248: ...248 Type 8098 FLOWave L English...
Страница 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Страница 304: ...304 Type 8098 FLOWave L English...
Страница 316: ...316 Type 8098 FLOWave L English...
Страница 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Страница 352: ...352 Appendix Type 8098 FLOWave L English...
Страница 353: ......
Страница 354: ...www burkert com...