Büchi Labortechnik AG
Help with faults | 10
Operation Manual Lyovapor™ L-300/L-300 Pro
109/118
Error
message
Possible cause
Solution
5570
Heatable shelf is disconnected.
u
Connect a functioning shelf at the
same connector position.
u
Contact BUCHI Customer Service.
Defective component (e.g. heat-
ing coil, cable, plug).
5571
Sample temperature sensor is
disconnected.
u
Connect a functioning PT-1000 at
the same connector position.
u
Contact BUCHI Customer Service.
Defective component (e.g. heat-
ing coil, cable, plug).
5704
System must be restarted after
changing the network settings.
u
Restart the instrument.