5 | Installation
Büchi Labortechnik AG
34/118
Operation Manual Lyovapor™ L-300/L-300 Pro
5.2.10 Fitting an air filter (option)
5
4
3
2
1
10
9
8
7
6
Fig. 20: Fitting an air filter
1
Air filter
2
Tube clip
3
Tubing
4
Tube clip
5
Gas connection for pressure regulat-
ing valve
6
Gas connection for venting valve
7
Tube clip
8
Tubing
9
Tube clip
10 Air filter
Fitting an air filter for the pressure regulating valve
Precondition:
R
The gas connection complies with the specified parameters. See Chapter 3.6
"Technical data", page 21
u
Switch the On/Off master switch to Off.
u
Fit the tube (3) onto the gas connection for the pressure regulating valve (5).
u
Fix the tube (3) with the tube clip (4).
u
Fit the air filter (1) onto the tube (3).
u
Fix the air filter (1) with the tube clip (2).
Fitting an air filter for the venting valve
Precondition:
R
The gas connection complies with the specified parameters. See Chapter 3.6
"Technical data", page 21
u
Switch the On/Off master switch to Off.
u
Fit the tube (8) onto the gas connection for the venting valve (6).
u
Fix the tube (8) with the tube clip (7).
u
Fit the air filter (10) onto the tube (8).
u
Fix the air filter (10) with the tube clip (9).