5 | Installation
Büchi Labortechnik AG
32/118
Operation Manual Lyovapor™ L-300/L-300 Pro
9
10
7
6
5
2
1
3
4
8
Fig. 18: Fitting the pressure sensors
1
Pressure sensor
2
Pressure sensor
3
Clamp, ISO-KF 16
4
Clamp, ISO-KF 16
5
Seal, ISO-KF 16
6
Connection for pressure sensor
7
Seal, ISO-KF 16
8
Connection for pressure sensor
9
Connection
10 Connection
Fitting pressure sensor one
u
Switch the On/Off master switch to Off.
u
Remove the shipping cap from the pressure sensor connection (6).
u
Fit the pressure sensor (1) and seal (5) to the pressure sensor connection (6) and
fix with the clamp (3).
u
Plug the connecting lead connection into the socket (10) on the pressure sensor
(1).
u
Plug the connecting lead connection into the socket marked Vacuum Sensor 1.
u
Select the sensor from the submenu
[Settings]
on the control panel.
Fitting pressure sensor two
u
Switch the On/Off master switch to Off.
u
Remove the blanking cap from the pressure sensor connection (8).
u
Fit the pressure sensor (2) and seal (7) to the pressure sensor connection (8) and
fix with the clamp (4).
u
Plug the connecting lead connection into the socket (9) on the pressure sensor
(2).
u
Plug the connecting lead connection into the socket marked Vacuum Sensor 2.
u
Select the sensor from the submenu
[Settings]
on the control panel.