Operating Pro control panel | 7
Operation Manual Lyovapor™ L-300/L-300 Pro
67/118
Setting
Option
Meaning
[Pressure limit]
Enter value
Absolute value for divergence
from the set pressure before the
sample protection function is ac-
tivated.
[Pressure duration]
Enter value
Sets the period of time that the
pressure is allowed to exceed
the pressure limit before the
sample protection function is ac-
tivated.
Editing a step
u
Use the function
[Right]
or
[Left]
on the function bar to navigate to the step that
you wish to edit.
u
Using the navigation control, navigate to the setting that you wish to change.
u
Tap the function
[Edit]
on the function bar.
u
Edit the setting as required.
u
Tap the function
[Save]
on the function bar.
ð
The setting is changed.
Add step
u
Use the function
[Right]
or
[Left]
on the function bar to navigate to the position
at which you wish to add a step.
u
Tap the function
[New]
on the function bar.
ð
The new step is created.
Delete step
u
Use the function
[Right]
or
[Left]
on the function bar to navigate to the step that
you wish to delete.
u
Tap the function
[Delete]
on the function bar.
u
When asked to confirm, press
[OK]
.
ð
The step is deleted.
7.6.7
Setting the phases of a method
NOTE
The settings in the Phase view affect all steps of a phase.
Navigation path
➔
Method
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The control panel highlights the selected method in green.
u
Tap the action
[[Phase]]
.
ð
The control panel shows the Phase view.
The following phases of a method are available: