Instrument Layout
daughter ion to be focused from the reflector together with the parent ion onto the
detector.
Thus post-acceleration by LIFT may be considered a segmented MALDI FAST
measurement using only one voltage segment to record a full daughter spectrum.
E
le
ct
ro
d
e
4
E
le
ct
ro
d
e
3
E
le
ct
ro
de
2
E
le
ct
ro
d
e
1
Ion Flight Direction
LIFT
Chamber
Pulsed
Acceleration
Stage
Acceleration
Stage
Ground
Ground
Ground
Ground
Phase 4
Phase 3
Phase 2
Phase 1
"Cell 1"
"Cell 2"
"Cell 3"
+ 19 kV
+ 19 kV
+ 19 kV
+ 19 kV
Figure 3-9 Potentials during a LIFT process on the related electrodes
ultraflex III User Manual, Version 1.0