○ Follow the 2-wire power cable to its nearby connector and disconnect the power.
○ Disconnect the tubing going to and from the pump
○ Unscrew the mounts for the pump and pull it out
○ Mount the new pump in its place and make sure it is secure
○ Connect the plumbing
○ Reconnect power
○ Re-enable the plumbing line on the Maintenance screen
● Replace Carriage
○ Fill the carriage with water and depressurize the lines
■ Go to the Maintenance Screen
■ Click on the Spray Head 1 card
■ Press the Remove Head button: This will depressurize the carriage and
fill it with water
■ Repeat for Spray Head 2
○ Disconnect the tubes going to the carriage
○ Disconnect the green electrical cable
○ Take out the screw that holds the carriage in place
○ Lift the carriage out of the Spray Chamber using the built-in handles
○ Place the new carriage in the carriage holder in the Spray Chamber
○ Reconnect the tubes and electrical connections
○ Pressurize the carriage by going to the Maintenance Screen → Spray Head 1→
pressurize. Repeat for Spray Head 2
■ Click the Begin Spray button on the Spray Head 1 tab
■ Verify that all nozzles have a continuous, full spray
■ Click Begin Spray button on the Spray Head 2 tab
■ Verify that all nozzles have a continuous, full spray
○ **If the carriage is being removed without replacement, disable pretreat from
spraying by changing the default fluid of both heads to Water:
■ Maintenance Screen
■ Spray Head 1 tab
■ Change the “Default:” fluid to Water
● Clicking next to “Default” (top right corner of tab window) will give
you a list of options. Choose Water
Addendum: Synergy Maintenance Page 6