CORRECTING COMMON PROBLEMS
If a solution is not readily apparent, use the table below to help identify and correct specific malfunctions. After you
correct the malfunction, verify that the generator successfully completes the self-test.
57
Ser vice Guide • Aaron 1250
Situation
Possible cause
Recommended action
Generator does not respond when
turned on
Disconnected power cord, faulty wall
receptacle, or faulty power cord
1.
2.
Check power cord connections (generator
and wall receptacle).
Connect the power cord to a functional wall
receptacle. If necessary, replace the power
cord.
Fuse drawer is open or fuses blown
1.
2.
Close the fuse drawer. If necessary, replace
the fuse(s).
If a problem persists, use a backup
generator.
Loose or disconnected internal cables
Check all internal connections.
Faulty power entry module or
connections
Check the power entry module and its cable
connections.
Faulty power switch
Replace the power switch.
Generator is on, but did not complete
the self-test.
An alarm condition exists.
Check the display for an error code. Note
the number and refer to Error Code list.
Loose or disconnected internal cables
Check and correct all internal connections.
Faulty power switch
Replace the power switch.
Main board malfunction
Replace the main board.
Display board malfunction
Replace the display board.
Activation and / or alarm tones do
not sound; speaker is malfunctioning.
Loose or damaged connection between
speaker board and main board
Check / connect all connections from the
speaker board to the main board.
Loose or disconnected cable between
main board and display board
Check / connect ribbon cable from the main
board to the display board.
Main board malfunction
Replace the main board.
Display board malfunction
Replace the display board.
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...