CONTROLS AND INDICATORS OVERVIEW
Users may control most Aaron 1250 functions from the front panel. Each Control is plainly marked and
colored on the front panel for quick reference. Volume control and a footswitch connector are located on
the rear panel.
Normal operations involve activating the generator with either a front connected handswitch OR a rear-
connected footswitch. The following components are the User Interface for the Aaron 1250 Electrosurgical
Generator.
Power Switch
The rocker ON / OFF switch on the lower left corner
allows the Aaron 1250 to be shut off when the unit is not
in use.
Membrane Function Switches
The front panel overlay contains 6 membrane function
switches (sometimes called matrix switches). There is a
membrane switch dedicated for each operational mode.
These switches toggle the unit between mode settings.
Power Control Knob
These rotary knobs allow you to select the desired
RF power level for all modes of operation. The Power
Control Knobs move at a graduated 1 watt per notch
(incrementally).
Watts Display A & B (Cut and Coag
)
These large Power Output Displays report the generator’s
output power setting from 1 to 120 watts in 1 watt
increments (at the rated load). During operation, the
numeral output of the display gives the surgeon an
indication of available generator power.
Visual LED Indictors
Mode LEDs indicate the mode setting.
The YELLOW indicators and controls indicate cutting
and blending operations. A yellow field LED indicates that
either a Cut or Blend mode is activated.
The BLUE indicators and controls indicate Coagulation,
Fulguration, and Bipolar operation. The blue field LED
indicates either Coagulation, Fulguration or Bipolar mode
is activated.
The Footswitch Control LED Indicator indicates which
mode the footswitch is presently in.
Monopolar footswitch control allows the user to activate
the monopolar mode when using footswitch controlled
accessories.
Bipolar footswitch control allows the user to activate the
bipolar mode.
17
Ser vice Guide • Aaron 1250
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...