RECOMMENDED EQUIPMENT FOR TROUBLESHOOTING
The following equipment enables you to troubleshoot and repair the Aaron 1250 Electrosurgical Generator.
• Digital multimeter with leads
• Electrosurgical analyzer or a true RMS voltmeter such as a Fluke 8920A
• Wideband current transformer such as a Pearson 4100
• Noninductive RF load resistors 200 W, 500 W, 800 W, 1000 W
• Oscilloscope (dual channel) at 100 MHz
• Oscilloscope probes, (2) 10X and 1000X
• Bovie/Aaron footswitch
• Bovie/Aaron handswitching pencil (single use or reusable)
• Standard technician’s tool kit
• Miscellaneous test leads and cables.
TROUBLESHOOTING THE AARON 1250
If the generator is not functioning properly, use the information in this section to perform the following activities:
• Identify and correct the malfunction.
• If an error code was displayed, take the appropriate action(s) to correct the error condition.
Inspecting the Generator
If the Aaron 1250 malfunctions, check for obvious conditions that may have caused the problem.
1. Check the generator for visible signs of physical damage.
2. Verify that all accessory cords are properly connected.
3. Check the power cord. Replace the power cord if you find exposed wires, cracks, frayed insulation, or a
damaged connector.
4. Open the fuse drawer and inspect the fuse housing and fuses for damage and corrosion.
5. Verify that the fuses are firmly seated. An internal component malfunction in the generator can damage the fuses.
6. You may need to replace the fuses if the generator fails the self-test or stops functioning. Refer to Fuse
Replacement in Section 6.
Inspecting the Receptacles
Equipment required:
• Footswitch
• Bipolar cable
• Monopolar instruments (handswitch and footswitch)
• Return electrode cable.
Procedure:
1.
Turn off the generator.
2.
Disconnect the power cord from the wall receptacle.
3.
Check the footswitch receptacle on the rear of the unit for obvious signs of obstruction and damage.
4.
Check for a secure fit by inserting the footswitch connector into footswitch receptacle.
If the footswitch receptacle is damaged, replace the footswitch connector assembly.
54
Bovie / Aaron Medical
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...