UNIT DESCRIPTION
The Aaron 1250 electrosurgical generator is a self-contained unit, consisting of the main enclosure and power cord.
The main components incorporated in the generator include:
•
FRONT PANEL COMPONENTS
Power switch; two dials for controlling power output; membrane switches for
selecting modes; receptacles for connecting electrosurgical accessories; and indicators that show the current settings
and patient return electrode status.
•
REAR PANEL COMPONENTS
Volume control; footswitch receptacle; power cable receptacle and fuse holder;
equipotential grounding lug; and remote accessory receptacle.
•
INTERNAL COMPONENTS
Display board; main board; pad sensing module; speaker board; and relay board.
SAFETY PRECAUTIONS WHEN REPAIRING THE GENERATOR
Before servicing the Aaron 1250 Generator it is important that you read, understand, and follow the instructions
supplied with it. Also, be familiar with any other equipment used to install, test, adjust, or repair this generator.
General Warnings, Cautions, and Notices
WARNINGS
Use the generator only if the self-test has been completed as described. Otherwise, inaccurate
power outputs may result.
The instrument receptacles on this generator are designed to accept only one instrument at a
time. Do not attempt to connect more than one instrument at a time into a given receptacle.
Doing so will cause simultaneous activation of the instruments.
CAUTIONS
Do not stack equipment on top of the generator or place the generator on top of electrical
equipment. These configurations are unstable and / or do not allow adequate cooling.
Provide as much distance as possible between the electrosurgical generator and other electronic
equipment (such as monitors). An activated electrosurgical generator may cause electrical
interference with them.
Do not turn the activation tone down to an inaudible level. The activation tone alerts the surgical
team when an accessory is active.
NOTICES
If required by local codes, connect the generator to the hospital equalization (grounding) connector
with an equipotential cable.
Connect the power cord to a wall receptacle having the correct voltage. Otherwise, product dam-
age may result.
Active Accessories
WARNINGS
Shock Hazard
– Do not connect wet accessories to the generator.
Shock Hazard
– Ensure that all accessories and adapters are correctly connected and that no
metal is exposed.
11
Ser vice Guide • Aaron 1250
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...