INSTALLATION AND PLACEMENT
Place the Aaron 1250 Electrosurgical Generator on any flat surface with a tilt angle of not more than 10 degrees.
The unit relies on natural convection cooling. Do not block the rear vents.
Ensure that air flows freely on all sides of the unit.
WARNING
Connect the power cord to a properly polarized and grounded power source with the frequency
and voltage characteristics that match those listed on the back of the unit.
FUNCTIONAL (OPERATIONAL) CHECKS
Upon initial installation of the unit, perform the following checks. Refer to the figures in Controls and Indicators for
the location of connectors and generator controls.
WARNING
At no time should you touch the active electrode or bipolar forceps. A serious burn could result.
How to Set Up and Start the Aaron 1250 Unit
1. Verify that the Power Switch is in the OFF position and that no accessories are connected to the unit.
2. Connect a hospital-grade power cable to the AC power cable receptacle on the back of the unit, then to a
properly grounded wall outlet.
3. Connect a two-pedal footswitch to the matching footswitch receptacle on the back of the unit. Use only Aaron
footswitches. Although other types of footswitches may fit, they may not be compatible with this Electrosurgical
Generator.
4. Do not connect a Patient Return Electrode to the front of the unit at this time.
5. Turn the unit on by switching the power switch to the (ON) position.
6. The correct startup for the unit is a quick flash of all indicators and a series of beeps. The unit will return to the
last mode and power setting used.
How to Check the Patient Return Electrode Alarm Function
1. Adjust the power settings for each mode (Cut, Blend, Coagulation, Fulguration, and Bipolar) to one watt.
2. Press the Cut Pedal of the footswitch.
3. Verify that an alarm sounds for three seconds, and then the Patient Return Electrode Sensing Alarm Indicator
Light illuminates. This indicates NO patient return electrode connected to the unit.
4. Verify that adjusting the volume control on the back of the unit (while the alarm is sounding) cannot change the
alarm’s sound level.
47
Ser vice Guide • Aaron 1250
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...