UNDERSTANDING ERROR CODES AND AUDIO TONES
The Aaron 1250 Electrosurgical Generator includes automatic, perpetual, self-diagnostics. If the diagnostics detect an
error, the system displays an error code, sounds an audible tone, and deactivates the output power.
Any errors detected will shut down the RF output power.
NOTICE
Internal firmware self-diagnostics continually monitor the unit’s operation to ensure proper and
safe performance.
Most error codes result from faults in accessories attached to the unit. The following table lists the error codes,
describes the error, and recommends actions to take to resolve the error.
56
Bovie / Aaron Medical
ERROR CODE
DESCRIPTION
RECOMMENDED ACTION
F1
(on the Cut /
Blend display)
F1
(on the
Coagulation /
Fulguration /
Bipolar display)
Handswitch or monopolar
footswitch cut pedal may be stuck
Handswitch or monopolar
footswitch coag pedal may
be stuck
1.
2.
3.
4.
Turn off, then turn on the generator. Do not press buttons or
accessory activation devices during the self-test.
If the alarm number reappears, disconnect all accessories.
Turn off, then turn on the generator again.
If the problem persists, replace the handpiece or footswitch
and repeat the restart.
If the alarm number reappears, record the number and call
the Aaron Medical Service Center.
F2
Cut and Coag buttons activated
simultaneously
(pencil or footswitch)
The unit does not allow simultaneous activation of the
cut and coagulation modes. Release either the cut or
coag button on the handpiece, or the cut or coag pedal
on the footswitch.
F3
Footswitch Cut or Coag Pedal
pressed while in Bipolar Foot
Control and the unit is not in
Bipolar Mode
The unit will not allow the Footswitch to activate the unit
if Bipolar Footcontrol is selected, but the Bipolar Mode
is not selected.
E4
Line voltage error
(Line voltage too high)
1.
2.
Turn the unit off.
Verify that the unit is connected to the correct line voltage.
E7
Internal temperature of the unit
exceeded limit
1.
2.
Turn unit off.
Allow unit to cool for 20 minutes.
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...