Monopolar Coag Curves
These measurements were taken using short (< 0.5 meter) leads.
Figure 3 – 5 Output power versus impedance for Coagulation mode
Figure 3 – 6 Peak voltage vs. power setting for Coagulation mode
31
Ser vice Guide • Aaron 1250
Load Resistance (ohms)
Output Power Setting (watts)
Open Circuit Peak V
oltage (volts)
Output Power (watts)