STANDARDS AND IEC CLASSIFICATIONS
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation failure because of the way in which
they are connected to the protective earth conductor.
Type BF Equipment (IEC 60601-1) / Defibrillator Proof
The Aaron 1250 Electrosurgical Generator provides a high degree of protection against electric shock,
particularly regarding allowable leakage currents. It is type BF equipment. Patient connections are isolated
from earth and resist the effects of defibrillator discharge.
Drip Proof (IEC 60601-2-2)
The generator enclosure is constructed so that liquid spillage in normal use does not wet electrical insulation or other
components which, when wet, are likely to affect adversely the safety of the generator.
Electromagnetic Interference
When other equipment is placed on or beneath an activated Bovie Aaron Medical electrosurgical generator, the
Aaron 1250 Electrosurgical Generator operates without interference. The generator minimizes electromagnetic inter-
ference to video equipment used in the operating room.
Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)
The Aaron 1250 Electrosurgical Generator complies with the appropriate IEC 60601-1-2 and IEC 60601-2-2
specifications regarding electromagnetic compatibility.
Voltage Transients (Emergency Generator Mains Transfer)
The Aaron 1250 Electrosurgical Generator operates in a safe manner when the transfer is made between line AC and
an emergency generator voltage source.
OUTPUT POWER CHARACTERISTICS
Maximum Output for Bipolar and Monopolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
* an indication of a waveform’s ability to coagulate bleeders without a cutting effect
28
Bovie / Aaron Medical
Mode
Output Power
Output Frequency
Repetition Rate
Vp-p max
Crest Factor*
(Rated Load)
Cut
120 W @ 500
Ω
357 kHz ± 50 kHz
N / A
2.5 KV
2.9 ± 20%
Blend
90 W @ 800
Ω
357 kHz ± 50 kHz
30 kHz ± 5 kHz
3.5 KV
3.3 ± 20%
Coagulation
80 W @ 1000
Ω
425 kHz – 494 kHz
57 kHz ± 5 kHz
4.5 KV
5.5 ± 20%
Fulguration
40 W @ 1000
Ω
410 kHz ± 50 kHz
25 kHz ± 5 kHz
6.5 KV
7.7 ± 20%
Bipolar
30 W @ 200
Ω
506 kHz – 570 kHz
32 kHz ± 5 kHz
2.0 KV
6.9 ± 20%
Содержание Aaron 1250
Страница 1: ...1 Service Guide Aaron 1250 SERVICE GUIDE...
Страница 8: ...8 Bovie Aaron Medical...
Страница 14: ...14 Bovie Aaron Medical...
Страница 16: ...FRONT PANEL Figure 2 1 Layout of controls indicators and receptacles on the front panel 16 Bovie Aaron Medical...
Страница 24: ...24 Bovie Aaron Medical...
Страница 35: ...Figure 3 13 Coagulation mode waveform Figure 3 14 Fulguration mode waveform 35 Service Guide Aaron 1250...
Страница 36: ...Figure 3 15 Bipolar mode waveform 36 Bovie Aaron Medical...
Страница 44: ...44 Bovie Aaron Medical...
Страница 50: ...50 Bovie Aaron Medical...
Страница 62: ...62 Bovie Aaron Medical...