17.2 Application
17.2.1 Basic application
The application specifies the basic characteristics. It is generally used to selectively control the
parameter list, i.e. many of the parameters which are not required are hidden. At the same time, some
functions in the device are (de-)activated, e.g. mass-relevant values are only calculated if mass has
been selected.
Selection:
ERW700
Mass
Linearization
UR06
Standard volume
Special
2-channel
Event cnt (event counters)
Batch
Energy
Subset. cnt (Subset counters)
Min/Max
ERW700
Activates the functions which are more relevant for energy distribution.
UR06
Activates the functions which are more relevant for the mineral oil industry.
2-channel
Activates functions which are relevant for applications with 2 more or less
independent units,
e.g. fuel (gas)/heating oil switchover, steam-condensate measurement
Energy
Activates functions which are required for energy calculation.
Mass
Activates functions which are required for mass calculation.
Standard volume
Activates functions which are required for standard volume calculation.
Event counters
Activates counters which are used to count events,
e.g. fault counters, overload counters, measured-value dependent counters
Special
Activates special functions.
Batch
Activates batch function
Min/Max
Min und Max-values for flow and power are stored
17.2.2 Language
Setting the dialog language
Selection:
German
, English, French, Spanish
17.3 Error external
Set the handling of external error contacts.
17.3.1 Selection
When set to "message" only an error message is generated, the reaction corresponds to the not
relevant for measurement errors. When set to "stop" the count is stopped or counted in fault counters.
Selection:
without
, message, stop
17.3.2 Input
Digital Input
Selection: DI1, DI1 inv, DI2, DI2 inv, DI3, DI3 inv, DI4, DI4 inv, DI5, DI5 inv, DI6, DI6 inv
17.4 Type label
17.4.1 Serial number
The factory-defined serial number is for information purposes only.