Universal Flow Computer UR06 

Operating Manual 

Page 44 of 55 

Subject to changes 

A-EN-08728-URRev.B 

Bopp & Reuther 

 

 

 

Messtechnik GmbH 

transmitter is > 19.5 mA. If a transmitter fails, calculation continues with the unaffected transmitter if it 
is in its measuring range. 

17.17.1  dp mode 

Operating mode for the differential pressure transmitter. 

Selection:  Default 

 

Calculation with default value (only for test purposes) 

ST  

 

Measurement with 1 transmitter 

ST rad

 

 

Measurement with 1 transmitter square rooted 

DT  

 

Measurement with 2 transmitters 

DT rad 

 

Measurement with 2 transmitters square rooted 

EWZ211   

Connection of a WZ200 transmitter 

DT31x.1   

Connection of a DT31x.1 with balancing 

DT31x.1 rad 

Connection of a DT31x.1 with balancing square root 

DT31x.1 + ST 

Connection of a DT31x.1 with bal ST 

DT31x.1 + ST rad  Connection of a DT31x.1 with bal ST square root 

17.17.2  dp default 

Default value for the differential pressure. Only for test purposes. 

Range: 

-3000.0 ... +3000.0 mbar dp  

17.17.3  dp averaging 

Factor for the average determination of the differential pressure. 

Range: 

1

 … 255 

17.17.4  dp offset 

Use the dp offset to correct static offset of the differential pressure caused by the transmitter or 
installation. The set value is subtracted from the measured value. 

Range: 

-10.0 mbar ... 

0.0

 … +10.0 mbar 

17.17.5  dp1A/1B end 

Maximum value (end value) of the dp transmitter at 20 mA. 

Range: 

1.0 mbar ... 10000.0 mbar 

17.17.6  dp1A/1B input 

Electrical assignment of the current input to the measured value. 

Range: 

I1, I2, I3, I4 

17.18 

Pressure 

17.18.1  Air pressure 

Air pressure for correcting the relative pressure sensors. 

Range: 

0.500 ... 

1.000

 … 1.200 bar 

17.18.2  Pressure Q correction 

Flow rate dependent pressure correction. 
P

k

 = P + pressure Q correction * density * Qb² 

Range: 

-1e37 ... +1e37 

17.18.3  Pressure mode 1 ... 3 

Operating mode for the pressure transmitter. The absolute pressure is corrected with –50% of the dp 
for the cross probe (CP). In the Q correction setting, the pressure is corrected in a flow rate dependent 
manner. 

Selection: 

without 

0/4..20 mA rel 

0/4..20 mA abs. KS 

default 

0/4..20 mA abs. 

0/4..20 mA abs. Qkorr. 

Содержание UR06

Страница 1: ...6 Speyer Phone 49 6232 657 0 Fax 49 6232 657 505 Internet http www bopp reuther de E Mail info bopp reuther de Dimensions weights and other technical details subject to change A EN 08728 UR Rev D 08 2...

Страница 2: ...s 10 7 2 1 Ambient temperature 10 7 2 2 Storage temperature 10 7 2 3 Climate class 10 7 2 4 Mechanical class 10 7 2 5 Degree of protection 10 7 2 6 Electromagnetic compatibility 10 8 Constructive Desi...

Страница 3: ...17 2 Application 34 17 2 1 Basic application 34 17 2 2 Language 34 17 3 Error external 34 17 3 1 Selection 34 17 3 2 Input 34 17 4 Type label 34 17 4 1 Serial number 34 17 4 2 TAG number 35 17 4 3 Or...

Страница 4: ...7 12 Density 40 17 12 1 Density determination 40 17 12 2 Density input 41 17 12 3 Density measuring parameter 0 1 41 17 12 4 Density default 41 17 12 5 Density end 41 17 12 6 Density start 41 17 12 7...

Страница 5: ...Pulse value digital output 1 3 7 47 17 22 Current output 47 17 22 1 Selection current output 1 2 6 47 17 22 2 Default current output 1 2 6 48 17 22 3 End current output 1 2 6 48 17 22 4 Start value cu...

Страница 6: ...nter 51 17 27 Balancing 51 17 27 1 PT100 balancing offset slope 1 2 51 17 27 2 PT500 1000 balancing offset slope 1 2 51 17 27 3 Current input balancing offset slope 1 2 6 51 17 27 4 Current output bal...

Страница 7: ...universal flow computer can result in hazardous situations Therefore always observe the warning signs 2 3 Operational safety UR06 must not be used in explosive areas The universal flow computer UR06...

Страница 8: ...put 4 1 Measured variable Electrical measured variables Current pulse frequency resistance contact status Physical measured variables Temperature pressure differential pressure volume flow mass flow d...

Страница 9: ...and 22 mA according to NAMUR NE43 Accuracy 0 02 of MV 0 002 mA Temperature influence 0 005 K Resolution 16 Bit Pulse Status Type Open collector passive galvanic isolated Frequency range 0 100 Hz Min p...

Страница 10: ...orage temperature 30 C to 70 C 7 2 3 Climate class In accordance with EN 1434 Class C 7 2 4 Mechanical class In accordance with EU directive 2004 22 EC Class M1 7 2 5 Degree of protection IP65 IEC 529...

Страница 11: ...ual Universal Flow Computer UR06 Bopp Reuther A EN 08728 URRev B Subject to changes Page 11 of 55 Messtechnik GmbH 8 Constructive Design 8 1 Model Dimensions 1 Standard plastic housing Front view Wall...

Страница 12: ...signment 9 1 Standard housing UR06 in the standard housing has 36 double row terminals connection to the display occurs via a western plug All terminal descriptions relate to hardware version HV3 Term...

Страница 13: ...ted active Iout 2 25 Current output galvanic isolated active DI 1 26 Digital input pulse frequency status DI 1 27 Digital input pulse frequency status DI 2 28 Digital input pulse frequency status DI 2...

Страница 14: ...galvanic isolated active Iout 1 6 Current output galvanic isolated active Iout 2 7 Current output galvanic isolated active Iout 2 8 Current output galvanic isolated active Module I IN Designation No F...

Страница 15: ...internal connected 4 M 5 M Bus M 6 Connected to terminal 5 M 7 M Bus M 8 Connected to terminal 7 Module RS485 Designation No Function Tx 1 Serial interface transmit Rx 2 Serial interface receive GND 3...

Страница 16: ...ent J4 J3 J2 J1 1 1 OC 1 4 1 MUS1 12 1 DI 1 26 2 2 OC 1 5 2 I1 13 2 DI 1 27 3 3 OC 2 6 3 GND 14 3 DI 2 28 4 4 OC 2 7 4 MUS2 30 4 DI 2 29 5 5 OC 3 8 5 I2 31 5 PT1 15 6 6 OC 3 9 6 GND 32 6 PT1 16 7 7 7...

Страница 17: ...e terminals X2 13 31 X3 14 32 10 2 Passive sensors Connection of sensors which do not have a power supply 2 wire transmitter Terminal Alternative terminals X1 12 30 X2 13 31 10 3 Temperature sensors C...

Страница 18: ...e digital sensors dual pulse Connection of digital sensors which provide a voltage signal The input has to be configured via the software Terminal Alternative terminals X1 26 1 5 X2 27 2 6 X3 28 3 7 X...

Страница 19: ...of 55 Messtechnik GmbH 10 7 2 channel passive digital sensors dual pulse Connection of digital sensors which have a passive output relay open collector NAMUR The input has to be configured via the so...

Страница 20: ...tput Connection 11 1 Current outputs Active current output Terminal Alternative terminals Alternative module terminals X1 22 24 5 7 X2 23 25 6 8 11 2 Digital outputs Optocoupler Terminal Alternative t...

Страница 21: ...ht is set to go off 10 minutes after the last keypress and the device switches to the standard display 12 2 1 Display of measured values UR06 1 Standard volume 1 0004711 987 m3 2 Volume 1 0123456 789...

Страница 22: ...ds Application Type label Display 12 2 3 Display of parameters Display of the parameter 0 M Bus Address 1 Access 0 Displayed when editing the parameter 0 M Bus Address 1 0 250 Header Parameter group P...

Страница 23: ...ing mode a numerical value decreases by 1 or you move one place down a value list Next Press this key to jump one level lower in the level structure Once you have reached the desired level press the k...

Страница 24: ...xample Depending on the configuration the number of values and their sequence may vary Current display UR06 Standard volume 1 0012345678 5 m3 Volume 1 0009833823 4 m3 1xMinus UR06 1 Standard volume 1...

Страница 25: ...Display 1xNext Parameters Password Level1 Password Level2 Password Level3 Password Level4 1xNext 0 Passwords Password Level1 Access 0 1xNext 0 Passwords Password Level1 0 9999 1xPlus 0 Passwords Passw...

Страница 26: ...xt Parameters Passwords Application Type label Display 13xMinus Parameters Temperature Thresholds Outputs Clock 1xNext Parameters Temperature Thresholds Outputs Clock 1xNext Clock Date Time Clock adju...

Страница 27: ...ubset Standard volume 2 event subset AUX 1 4 AUX 2 4 AUX 3 4 AUX 4 4 Instant values Power 1 Mass flow rate 1 Volume flow rate 1 Standard volume flow rate 1 Temperature 1 Temperature 1 subset 1 Tempera...

Страница 28: ...w computer was not connected to the power supply Measuring hours Time during which the flow rate was above the leak flow rate Error hours Time during which an error occurred Saturated steam hours Time...

Страница 29: ...emperature 1 Date Time 13 07 10 14 02 00 Max 48 7 C Day storage 1 Temperature 1 Date Time 03 01 10 03 12 00 Min 0 97 C 15 3 Parameter log Changes to the parameters are saved together with the old valu...

Страница 30: ...6 period log Batch No Values 14 Checksum 8B39h log memory 2 296 start time 03 05 12 15 15 00 stop time 03 05 12 15 30 00 log memory 2 296 Error 00000000b 00h Status 00000000b 00h log memory 2 296 1 En...

Страница 31: ...menu is arranged as a tree structure To illustrate the structure certain levels are displayed below as a mind map Branches without a dot are end branches branches with a dot at the end continue to the...

Страница 32: ...Universal Flow Computer UR06 Operating Manual Page 32 of 55 Subject to changes A EN 08728 URRev B Bopp Reuther Messtechnik GmbH 16 1 3 Level 2 and 3...

Страница 33: ...ve password level must be released first via the current valid password Password level L0 no password Only the communication parameters for M Bus and Modbus can be altered here Password level L1 Acces...

Страница 34: ...witchover steam condensate measurement Energy Activates functions which are required for energy calculation Mass Activates functions which are required for mass calculation Standard volume Activates f...

Страница 35: ...number Assignment of list values to a line in the display If 0 is entered the value is not displayed If the same line number is assigned to several values they are displayed alternately The available...

Страница 36: ...t is m3 h If the transmitter is a mass transmitter or an orifice the unit is kg h Range 0 001 m3 h 1e9 m3 h or kg h 17 6 4 Pipe diameter Internal diameter of the supply pipe The input of the correct d...

Страница 37: ...w the leak flow rate Selection Not count count 17 6 12 Calibration threshold The calibration threshold is set above the leak flow rate for measurements as permitted by the German Board of Weights and...

Страница 38: ...r of the bore Range 3 1500 mm 17 7 5 TC pipe Temperature coefficient of the supply pipe e g tc V2A 16 E 6 1 K Range 1 100 E 6 1 K 17 7 6 TC bore Temperature coefficient of the bore Range 1 100 E 6 1 K...

Страница 39: ...ss O C low pass 17 10 Fluid 17 10 1 Fluid Selection of the fluid The density enthalpy and dynamic viscosity for standard media are calculated according to internally defined methods The polynominal co...

Страница 40: ...essure a vapor pressure and volume fault is reported Selection No Yes 17 10 8 Pressure reserve The switching point of the vapor pressure monitor can be controlled via the pressure reserve Steam switch...

Страница 41: ...sity start Minimum value start value of the density transmitter at 0 4 mA Range 0 0 kg m3 10 000 0 kg m3 17 12 7 Temperature Allocation of a measured temperature to the density meter Selection Tempera...

Страница 42: ...to the measured value Range I1 I2 I3 I4 17 14 3 Default Default value for the concentration Range 0 100 17 14 4 End value Value of the concentration at 20 mA Range 0 100 17 14 5 Start value Value of t...

Страница 43: ...sity SO3 P3 Density SO4 P4 17 16 2 Density 15 C Input of the density at 15 C An error is displayed if the entered density does not match the oil group Range 100 2 000 0 kg m3 17 16 3 Compressibility m...

Страница 44: ...purposes Range 3000 0 3000 0 mbar dp 17 17 3 dp averaging Factor for the average determination of the differential pressure Range 1 255 17 17 4 dp offset Use the dp offset to correct static offset of...

Страница 45: ...e level and the transmitter center a value of 0 45 bar has to be set 17 18 8 Pressure input Electrical assignment of the current input to the measured value Range I1 I2 I3 I4 17 19 Temperature 17 19 1...

Страница 46: ...0 s 17 20 3 Upper Lower limit monitoring The threshold of the allocated measured variable whose violation generates an error message once the grace period has expired The measured value then changes t...

Страница 47: ...d to the setting in Pulse mode Selection DDS pulse 17 21 3 Digital output 1 3 7 Selection of the digital output operating mode Sums or differences can be output via AUX 1 4 The function is defined und...

Страница 48: ...6 Damping factor for the current output Tau n 125 ms Range n 1 255 17 22 6 Error behavior current output 1 2 6 Behavior of the current output if an error occurs Depending on the setting the current ou...

Страница 49: ...n time the min max of this value is then tested and if necessary saved Range 0 15 255 min 17 24 5 Data log configuration error comes goes Several log errors can be chosen to trigger a log event Select...

Страница 50: ...asic units for internal calculation are l kg Wh A decade overflow of the displayed counters occurs for each 1012 of the basic unit A decade overflow of the transferred counters occurs for each 109 of...

Страница 51: ...r status is displayed and frozen Use Next to accept the possibly changed value in the counter Range Counter range 17 27 Balancing 17 27 1 PT100 balancing offset slope 1 2 Calibration values for the PT...

Страница 52: ...cause of the error the messages are automatically deleted Only pulse errors have to be acknowledged If all the errors have been eliminated a message appears stating that new errors have been listed i...

Страница 53: ...3 6mA No pressure transmitter connected or no default value for gaseous media Measured value fails or calculation is carried out with default values E37 E69 Conf density 1 Conf density 2 Error in the...

Страница 54: ...wrong slot MODUL0 Current 3 and 4 are set to 0 mA E139 E140 E141 M Iin34 malfun M Iout34 malfun M Iout56 malfun Communication with module faulty E142 E143 M Iout34 config M Iout56 config Incorrect co...

Страница 55: ...ature Degree Celsius C 1 C 1 C Kelvin K TK 273 15 C TC 273 15 K Degree Fahrenheit F TF 32 5 9 C TC 1 8 32 F Degree Rankin Ra TRa 5 9 273 15 C TC 1 8 491 67 Ra Energy Watt hours Wh 1000 kWh 0 001 Wh Ki...

Отзывы: