Operating Manual
Universal Flow Computer UR06
Bopp & Reuther
A-EN-08728-URRev.B
Subject to changes
Page 51 of 55
Messtechnik GmbH
Volume event subset
Mass event
Energy event subset
Standard volume
Masse subset
Standard volume event
Mass event subset
17.26.4 Clear counters
All the counters are reset to 0. Alternatively, only counter groups can be cleared.
Commands:
Clear all counters
(All counters are cleared)
Clear all counters1
(All counters of group 1, primary transmitter are cleared)
Clear all counters2
(All counters of group 2, secondary transmitter are cleared)
Clear subsets1
(All subsets of group 1, primary transmitter are cleared)
Clear subsets 2
(All subsets of group 2, secondary transmitter are cleared)
Clear event 1
(All event counters of group 1, primary transmitter are cleared)
Clear event 2
(All event counters of group 1, secondary transmitter are cleared)
17.26.5 Set value counter
Changing and saving the new value.
(When selecting the parameter, the current counter status is displayed and frozen. Use “Next” to
accept the possibly changed value in the counter.)
Range:
Counter range
17.27
Balancing
17.27.1 PT100 balancing offset / slope 1 ... 2
Calibration values for the PT100 input.
Range:
According to the parameter
17.27.2 PT500/1000 balancing offset / slope 1 ... 2
Calibration values for the PT500/1000 input.
Range:
According to the parameter
17.27.3 Current input balancing offset / slope 1 ... 2 (6)
Calibration values for the current input.
Range:
According to the parameter
17.27.4 Current output balancing offset / slope 1 ... 2 (6)
Calibration values for the current output.
Range:
According to the parameter
18 Connection of the Universal Flow Computer
Energy supply:
Supply voltage:
180 V ... 264 V AC,
18 V ... 36 V DC (optional)
Cable gland:
1x M20, 3x M16 and 5x M12
Cable diameter:
3.5 to 12 mm
Terminals:
Externally tensioned plug-in terminals
Wire cross section:
0.5 … 1.5 mm² rigid and flexible
Modules 0.4 … 0.8 mm² rigid and flexible
Cable type:
We recommend LiYY (stranded pair, without shielding)
Cable length:
For sensors in accordance with EU directive 2014/32/EU
Annex MI-004 to 100 m. For outputs and other installations up to
500 m, depends on the ambient conditions.