E11
EEprom Para.
Parameter after device restart outside
the permissible range
Default value is used for the
affected parameter.
E12
EEprom count.
Incorrect check sum of the counters
saved in the EEPROM
E13
Pulses>max
Buffer overflow. Pulse value or
minimum pulse width has been set
incorrectly.
Output pulses are lost. No
influence on the displayed
counters.
E14
Mathematics
Incorrect configuration, incorrect
parameters
No re-evaluation
E15
Timeout LCD
Communication between the basic
device and the display interrupted
No influence on the re-
evaluation
E16
CRC EEPROM
Log (EEPROM) defective or not
installed
E17
Time setting
Clock defective
Incorrect time, tariff values
incorrect. No influence on the
re-evaluation
E33
E65
E97
Config. temp. 1
Config. temp. 2
Config. temp. 3
Error in the configuration of the
temperature input.
E34
E66
E98
Temperature 1
Temperature 2
Temperature 3
Temperature measurement faulty.
Incorrect configuration, short circuit,
cable break.
E35
E67
E99
Config. pressure 1
Config. pressure 2
Config. pressure 3
Error in the configuration of the
pressure input.
E36
E68
E100
Pressure 1
Pressure 2
Pressure 3
Measured variable outside the
measuring range. Wiring error.
Pressure transmitter is selected but not
connected or the current is less than
3.6mA.
No pressure transmitter connected or
no default value for gaseous media.
Measured value fails or
calculation is carried out with
default values
E37
E69
Conf. density 1
Conf. density 2
Error in the configuration of the density
measurement.
E38
E70
Density 1
Density 2
Density measurement faulty or outside
the specified area for the fluid.
Default value is used for
calculations.
E39
E71
Conf. dp 1
Conf. dp 2
Error in the configuration of the
differential pressure input.
dp=0
E40
E41
E42
E72
E73
E74
dp 1
dp 1A
dp 1B
dp 2
dp 2A
dp 2B
Measured variable differential pressure
outside the measuring range. Wiring
error. Differential pressure transmitter
is selected but not connected or the
current is less than 3.6mA.
dp=0
E43
E75
Conf. pri. transm.
Conf.sec. transm.
Error in the configuration of the primary
transmitter / secondary transmitter.
Flow rates are set to 0
E44
E76
Primary transmitter
Secondary
transmitter
Mass/Volume flow is set to 4-20mA for
the transmitter and the current falls
below 3.6 mA.
Flow rates are set to 0
E45
E77
Base density 1
Base density 2
E46
E78
Saturated steam 1
Saturated steam 2
Saturated steam monitor active and
measured values outside the saturated
steam line.
Saturated steam parameters
are used for the further
calculation.
E47
E79
GERG 1
GERG 2
Incorrect GERG parameters.
Calculation with constant
compressibility.
E49
E81
Comp pulse 1
Comp pulse 2
Number of permitted error pulses
exceeded. (only dual pulse)
E50
E82
Zero balance 1
Zero balance 2
Zero balance faulty (only EWZ211)
dp=0
E51
Volume 1
Measured variable fault fort he volume
No re-evaluation