Boonton 4540 Series RF Power Meter
Operation
4-38
Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency
Description:
Set the RF frequency for the current measurement. The appropriate frequency calfactor
from the sensor’s EEPROM table will be interpolated and applied automatically.
Application of this calfactor cancels out the effect of variations in the flatness of the
sensor’s frequency response. If an explicit calfactor has been set, either manually or via
the
SENSe:CORRection:CALFactor
command, entering a new frequency will override
this calfactor and use only the ―automat
ic‖
frequency calfactor.
Menu Type:
Numeric value.
Range:
0.001 GHz to 110.00 GHz (sensor and video bandwidth dependant)
SCPI Command:
SENSe[1|2]:AVERage <numeric_value>
Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms
Description:
Selects the alarms menu for the selected channel.
Menu Type:
Menu Select.
Содержание 4541
Страница 14: ...Boonton 4540 Series RF Power Meter Contents xiv Limited Warranty 11 1 ...
Страница 44: ...Boonton 4540 Series RF Power Meter Getting Started 3 16 Figure 3 5 Menus for Sensor Calibration Procedure ...
Страница 122: ...Boonton 4540 Series RF Power Meter Operation 4 74 FFF Fig 4 62 ...
Страница 222: ...Boonton 4540 Series RF Power Meter Remote Operation 5 88 This page intentionally left blank ...
Страница 250: ...Boonton 4540 Series RF Power Meter Appendix A 7 4 This page intentionally left blank ...
Страница 254: ...Boonton 4540 Series RF Power Meter Appendix A 8 8 This page intentionally left blank ...
Страница 258: ...Boonton 4540 Series RF Power Meter Appendix B 9 4 This page intentionally left blank ...